Home About Product Workshop Industry News Contact Ch En Sp Ar

Use and Maintenance of Centrifuges

Date:2020/6/28 13:28:05 Hits:1629

1. Before operating the centrifuge, the power should be cut off and the centrifuge brake should be released. You can manually rotate the drum to see if there is any brake bite.
2. Check other parts for looseness and abnormal conditions.
3. Turn on the power and drive clockwise to start (usually it takes about 40-60 seconds from a stationary state to normal operation). Usually, each equipment must run empty for about 3 hours after arriving at the factory, and can operate without any abnormalities.
4. The materials should be placed evenly as much as possible, and must be operated by dedicated personnel. The capacity should not exceed the designated quantity.
5. It is strictly prohibited to operate the machine at excessive speed to avoid affecting its service life.
6. After the machine starts, if there are any abnormal situations, it must be stopped for inspection, and if necessary, it must be disassembled, washed, and repaired.
7. The centrifuge operates at high speed, so it is not allowed to touch the drum with your body to prevent accidents.
8.The mesh size of the filter cloth should be determined based on the size of the solid particles of the separated material, otherwise it will affect the separation effect. In addition, the filter cloth should be installed properly.
9. The sealing ring is embedded in the sealing groove of the drum to prevent materials from entering.
10. To ensure the normal operation of the centrifuge, please refuel and maintain the rotating parts every 6 months. At the same time, check the lubrication condition of the bearings for any signs of wear; Check if there is any wear on the components in the braking device, and replace them in severe cases; Check for oil leakage in the bearing cover.
11. After the machine is used, it should be cleaned properly to keep it tidy.
12. Do not separate non corrosion-resistant centrifuges from highly corrosive materials; In addition, strictly follow the equipment requirements and regulations for operation. Non explosion-proof centrifuges must not be used in flammable or explosive situations.

Next:Nothing

Zhangjiagang Yongjia Purification Equipment Co., Ltd

Landline: 0086-512-58960600

phone: 86-18013069378

fax: 0512-58960680

E-mail: yjgfx@yjlxj.com

address:Leyu Economic Development Zone, Zhangjiagang City, Jiangsu Province

Copyright Copyright:Zhangjiagang Yongjia Purification Equipment Co., Ltd 备案号:Su ICP Bei No. 10209999